Please use this identifier to cite or link to this item:
http://earsiv.odu.edu.tr:8080/xmlui/handle/11489/4573
Title: | Cortical Bone Strains Around Straight and Angulated Immediate Orthodontic Microimplants: A Pilot Study |
Authors: | Cehreli, Secil Yilmaz, Alev Arman-Ozcirpici, Ayca Ordu Üniversitesi 0000-0002-9250-3689 |
Keywords: | bone, strain, mini-implant, torque application, orthodontic loading LOADED IMPLANTS, MATHEMATICAL-DESCRIPTION, CELLULAR ACCOMMODATION, PLACEMENT TORQUE, MINI-IMPLANTS, STABILITY, ANCHORAGE, ARCHITECTURE, PRINCIPLE, THICKNESS |
Issue Date: | 2013 |
Publisher: | LIPPINCOTT WILLIAMS & WILKINS-PHILADELPHIA |
Citation: | Cehreli, S., Yilmaz, A., Arman-Ozcirpici, A. (2013). Cortical Bone Strains Around Straight and Angulated Immediate Orthodontic Microimplants: A Pilot Study. Implant Dent., 22(2), 133-137. https://doi.org/10.1097/ID.0b013e31827b96c2 |
Abstract: | Purpose: To measure strains around orthodontic implants upon torque tightening and loading and to assess correlations between factors influencing primary stability. Materials and Methods: Self-drilling implants were placed into bovine iliac crest blocks after CT assessments. Upon bonding of strain gauges on bone adjacent to the implants, strain measurements were performed using a data acquisition system during torque tightening and 250 g orthodontic force application by elastic chains. Results: The torque required to place straight implants (12.16 N.cm) was higher than 30- to 40-degree angulated implants (9.31 N.cm) (P < 0.05). Cortical bone strain amplitudes of both implant placements were comparable (P > 0.05). Strains during torque tightening of straight (196 me) and tilted (114 mu is an element of) implants were higher than those obtained during orthodontic loading (20-30 mu is an element of). Despite the positive and direct relationship found between torque and torque strain output, strong correlations between other parameters could not be detected. Conclusion: Vertically aligned and 30- to 40-degree angulated immediate orthodontic microimplants are associated with low amplitude strains upon torque tightening and orthodontic loading. (Implant Dent 2013;22:133-137) |
Description: | WoS Categories: Dentistry, Oral Surgery & Medicine Web of Science Index: Science Citation Index Expanded (SCI-EXPANDED) Research Areas: Dentistry, Oral Surgery & Medicine |
URI: | http://dx.doi.org/10.1097/ID.0b013e31827b96c2 https://www.webofscience.com/wos/woscc/full-record/WOS:000316309900006 http://earsiv.odu.edu.tr:8080/xmlui/handle/11489/4573 |
ISSN: | 1056-6163 |
Appears in Collections: | Ortodonti |
Files in This Item:
There are no files associated with this item.
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.