Please use this identifier to cite or link to this item: http://earsiv.odu.edu.tr:8080/xmlui/handle/11489/1539
Title: Biomechanical Comparison of Transoral and Transbuccal Lateral Cortical Plate Fixation for the Management of Mandibular Angle Fractures
Authors: Ayranci, F.
Dayi, E.
Ghahramanzadehasl, H.
Karagol, S.
Omezli, M. M.
Polat, M. E.
Ordu Üniversitesi
0000-0001-7126-5696
0000-0002-5750-1143
0000-0002-6606-6593
Keywords: Miniplate fixation; stability; transbuccal
MINIPLATE; OSTEOSYNTHESIS; TITANIUM; CONDYLE; TRIAL; SCREW
Issue Date: 2017
Publisher: MEDKNOW PUBLICATIONS & MEDIA PVT LTD, B-9, KANARA BUSINESS CENTRE, OFF LINK RD, GHAKTOPAR-E, MUMBAI, 400075, INDIA
Abstract: Objectives: The purpose of this experimental study was to compare the biomechanical behaviors of two different types of osteosynthesis that are used in the treatment of mandibular angle fractures. Materials and Methods: Twenty synthetic polyurethane human mandible replicas, with medullar and cortical portions, were used in this study. These polyurethane hemimandibles were randomly divided into two groups (n = 10). The transbuccal group (Group A) was fixed with 7 mm long self-tapping 2.0 mm titanium screws at 85 degrees to the reference line and the transoral group (Group B) was fixed with the same screws at 15 degrees to the reference line. All testings were performed on a servo-hydraulic testing machine. The data were transmitted directly from the load cell to a computer, which showed the emergent results of the material characteristics under resisted forces as a graphic containing force and displacement. The peak point loading and displacement for each subject were measured. Results: The comparison between the groups was analyzed with an independent-samples t-test, and P < 0.05 was considered to be significant. The results show that there were no significant differences between the groups for the peak loads and displacement values at the peak loads. Conclusion: The results of this experimental study demonstrated that there were no significant differences between the transbuccal and transoral methods in terms of fixation stability. In other words, the screw position and angle seemed to no have influence on the fixation stability in single miniplate treatments of a mandibular angle fracture.
URI: http://doi.org/10.4103/1119-3077.187319
http://earsiv.odu.edu.tr:8080/xmlui/handle/11489/1539
Appears in Collections:Ağız, Diş ve Çene Cerrahisi

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.